Effect of Implant Diameter and Cortical Bone Thickness on Biomechanical Performance of Short Dental Implant-Supported Distal Cantilever: A Finite Element Study

Authors

DOI:

https://doi.org/10.4186/ej.2021.25.2.175 Full article

Abstract

The present study aimed to assess the effect of implant diameter and cortical bone thickness on the biomechanical performance of short dental implant-supported distal cantilever placed in posterior atrophic maxilla by finite element method. There were 6 stimulating models in this study, which consisted of a bone block and short dental implant of 7 mm in length, 3 differences of diameter (5.0, 5.5, and 6.0 mm) and 2 variations of cortical bone thickness (thickness of 0.5 and 1.0 mm). All models were applied with 200 N of force with 30° inclination to lingual direction. Von Mises stress and strains were used to evaluate the biomechanical performance of the short dental implant and surrounding bone. Results showed that increasing of implant diameter led to decrease in stress level on neck of dental implants and strains in surrounding cortical bone. However, the implant diameter equal or smaller than 5.0 mm should not be used due to high stress level with could be a cause of implant failure. In addition, clinicians should be aware of placing implant with inadequate bone stock which lead to surrounding bone resorption.

Keywords:

dental implant, biomechanics, distal cantilever, posterior maxilla

Affiliations

  • Suthep Muangsisied Thammasat University
  • Nattapon Chantarapanich Kasetsart University Sriracha
  • Melvin Stanley Veerasakul Kasetsart University Sriracha
  • Samroeng Inglam Thammasat University

Corresponding author: Samroeng Inglam, isamroeng@gmail.com

1036 942

Author Biographies

  • Faculty of Dentistry, Thammasat University, Khlong Luang, Pathumthani 12120, Thailand

  • Department of Mechanical Engineering, Faculty of Engineering at Sriracha, Kasetsart University Sriracha, Chonburi 20230, Thailand

  • Department of Mechanical Engineering, Faculty of Engineering at Sriracha, Kasetsart University Sriracha, Chonburi 20230, Thailand

  • Faculty of Dentistry, Thammasat University, Khlong Luang, Pathumthani 12120, Thailand

Downloads

How to Cite

[1]
S. Muangsisied, N. Chantarapanich, M. S. Veerasakul, and S. Inglam, “Effect of Implant Diameter and Cortical Bone Thickness on Biomechanical Performance of Short Dental Implant-Supported Distal Cantilever: A Finite Element Study”, Eng. J., vol. 25, no. 2, pp. 175–182, Feb. 2021, doi: 10.4186/ej.2021.25.2.175.

Citations

Published

2021-02-28

Issue

Section

Special Section for ICEIT 2020